Cooling bath thermostats with insulated cooling baths are an economic solution for thermoregulation of objects directly in the thermostat bath. The units are therefore suitable for a number of typical heating and cooling tasks such as sample temperature control, materials testing, quality control, analyses, etc. Cooling circulator operates with natural refrigerants and are therefore offer an environmentally friendly solution to temperature control. In addition, cooling circulators are fitted with automatic refrigeration capacity adjustment, reducing energy consumption and waste heat down to the absolute essential. With low level protection and adjustable over heating protection the units fulfil the requirements of Safety Class III/FL according to DIN 12876. The thermostats can also be used for externally attached applications using the pump connections at the rear.
General information & temperature controller
Temperature range |
-25...200 °C |
Temperature stability |
±0.02 °C |
Controller |
Pilot ONE  (E-grade "Basic") |
Safety class |
III / FL Â |
Protection class |
IP20 Â |
Dimensions (W x D x H) |
210 x 400 x 546 mm  |  8.27 x 15.75 x 21.5 inches |
Weight |
25 kg  |  55.12 pounds |
Â
Heating / cooling capacity
Heating capacity |
1,5Â kW |
Cooling capacity |
20 |
0 |
-10 |
-20 |
 °C |
0,26 |
0,21 |
0,15 |
0,05 |
 kW |
|
Cooling machine |
air-cooled  |
Refrigerant (ASHRAE, GHS) |
R290 (A3, H220) (0,047 kg) |
Â
Circulation pump
Pressure pump max. |
27 lpm ; 0.7 bar  |  7.13 gpm ; 10.15 psi |
Suction pump max. |
22 lpm ; 0.4 bar  |  5.81 gpm ; 5.8 psi |
Pump connection |
M16x1 AG male  |
Pump speed adjustable |
Yes |
Â
Connections
Pt100 connection |
Pt100 Â |
Digital interface |
RS232, Ethernet, USB Device, USB Host |
Â
Operating data
Permissible ambient temp. |
5...40 °C |
Bath opening W x D / BD |
140 x 120 / 150 mm  |  5.51 x 4.72 / 5.91 inches |
Bath volume |
4.5 Liters  |  1.19 gal |
Power supply |
110-120V 1~ 50/60Hz   |
Current max. |
15 + 3,2Â A |
€{{amount_with_comma_separator}}